Measure Frequencies from 1 to 20 000 Hz.
Measure Low-Frequency Sound and Noise with a Single Unit. |
![]() |
With the auto store function included as standard, as well as a timer function and external power supply support, the NL-62 is ideal for continuous measurement. Designed for intuitive ease of use, there is no more need to consult the manual during a measurement. The large 3-inch color screen is bright and easy to read. Sudden rainfall is also no problem, thanks to the water-resistant construction. Using the optional octave and 1/3 octave band real-time analysis program NX-62RT (under development), the unit can even operate as a frequency analyzer. The High-Precision Sound Level Meter NL-62 supports all your measurement needs. | |
Sound Level Meter class1 (With low- frequency sound measurement function) NL-62 |
Features
Large color LCD screen
Three-inch LCD screen with a touch panel High resolution screen is easy to see indoors or outdoors and even in the dark.
![]() |
![]() |
No paper manual is needed.
User instructions and a help function can be easily accessed on the device.
![]() |
![]() |
![]() |
Measurement Display (Level-Time graph) |
Measurement Display (low-frequency sound) |
Parameter Screen |
![]() |
![]() |
Menu screen |
Help screen |
Water-resistant (Except for the microphone) |
Use of rechargeable batteries |
Guaranteed water-resistant to at least level IP54 (resistant to spraying water).Helps reduce failures caused by sudden rain showers. ![]() * Mounting the All-weather windscreen or rainproof windscreen helps raise the water-resistant performance of the entire unit, so that the microphone will meet IPX3 specifications. |
In these new models it is possible to use rechargeable batteries which make these meters environmentally-friendly. 16 hour continuous measurement is possible (when using eneloop® or dry alkaline batteries). ![]() |
? Please use the dedicated charger to charged eneloop® batteries.
? When using eneloop batteries, please read the eneloop® battery instruction manual. ? eneloop® is a registered trademark of Panasonic group. |
Continuous detailed measurements for one month
This meter can be used to conduct long-term measurements, such as environmental measurements.(If an AC adapter is used)

Example of detailed recording |
If the Lp is measured at 100 ms intervals and the Leq is simultaneously measured at 10 min intervals over a 24 h period, the total size of accumulated data is approximately 74 MB (reference value) |
Functionality can be extended by a range of options
Add long-term data recording capability and frequency analysis function
![]() |
![]() |
1/3 octave band analysis screen (low range) |
FFT analysis screen (x40) |

Data management screen of AS-60 software
Program function list
Program function list
Auto store function
This function enables continuous measurement in Lp mode (instantaneous SPL) and Leq mode (equivalent continuous SPL) to be conducted simultaneously.
Total measuring time of Auto store function |
Up to 1 000 h Equipped with a timer function |
Lp mode (instantaneous SPL) and Leq mode (equivalent continuous SPL) concept

Comparator function
This function turns on when the open collector output exceeds the set value (max. applied voltage
24 V, max. current 60 mA, allowable dissipation 300 mW).

Continuous data output function
This function enables the continuous acquisition of instantaneous values and processed values during both USB and RS-232C communication.
This is a convenient function for users who can design their own control programs, where data has to be transferred continuously from the sound level meter to the computer.
This is a convenient function for users who can design their own control programs, where data has to be transferred continuously from the sound level meter to the computer.
Optional program function list
Optional program function list
Octave, 1/3 octave real-time analysis program
NX-62RT |
![]() The 512 MB SD card can be used as a memory card after installing the program. |
By adding the NX-62RT program to the NL-62, octave band and 1/3 octave band real-time analysis can be realized. Saved analysis results can be loaded and shown in an overlay graph display together with current analysis data. NC curve graph display and NC value calculation/display are also possible.
![]() |
![]() |
![]() |
1/3 octave band analysis screen (low range) |
1/3 octave band analysis screen (high range) |
Overlay analysis screen |
![]() |
![]() |
1/3 octave band analysis screen (combined bands) |
Measurement screen (Level-Time graph) |
Waveform recording program
NX-42WR |
![]() The 2 GB SD card can be used as a memory card after installing the program. |
This function enables users to record sounds and processing sound to levels simultaneously. Recorded data can be played on computer and used for frequency analysis.(Uncompressed waveform WAVE file) Sampling at 48 kHz, 24 kHz, 12 kHz, Selection of 24 bit or 16 bit |
Maximum recording time (16 bit)
|
Reverberation Time Measurement Program
NX-42RV |
![]() The 512 MB SD card can be used as a memory card after installing the program. |
By adding the NX-42RV program to the NL-52/42, reverberation time measurements can be performed. The measurement method is the interrupted noise method. This program allows storage of reverberation time decay curves, T20/T30 calculation, Txx calculation (reverberation time calculation based on a user-defined interval) and averaged reverberation time results displayed on the SLM screen.
![]() |
![]() |
![]() |
![]() |
Measuring screen (graph) |
Measuring screen (numeric) |
Reverberation time decay curve screen |
Result screen (T20/T30/Txx) |
FFT analysis program
NX-42FT |
![]() The 512 MB SD card can be used as a memory card after installing the program. |
By adding the NX-42FT program to the NL-62, FFT analysis can be performed. The analysis frequency range is 20 kHz, with 8 000 spectrum lines (200 displayed). Saved analysis results can be loaded and shown in an overlay graph display together with current analysis data. Maximum zoom ratio is x40, and the top list screen can show up to 20 lines.
![]() |
![]() |
![]() |
Analysis screen (x1) |
Analysis screen (x40) |
Overlay analysis screen |
![]() |
![]() |
Linear average screen |
Top list screen |
System construction
System construction

Peripheral devices
Peripheral devices
All-weather |
Rain-protection |
Sound calibrator |
Pistonphone |
||||||||
windscreen WS-15
![]() This windscreen is designed for outdoor installations. It helps to reduce wind noise and is equipped with rainproof features that satisfy the IPX3 water-resistant specifications. It is used with a microphone extension cable.
(Mounting adapter WS15006 required separately)(For All-weather windscreen WS-15, use of ST-81 is recommended.) |
windscreen WS-16
![]() This screen protects the microphone against rain for a short period of time.The rainproof performance of this windscreen is designed to satisfy the IPX3 water-resistant specifications.
|
NC-75
![]()
This Sound calibrator conforms to IEC 60942:2020 class 1(JIS C 1515:2020 class 1), providing a level of performance sufficient for calibrating the precision sound level meter.
Specifications
|
NC-72B
![]()
This Pistonphone conforms to IEC 60942:2017 class LS/M,class 1/M(JIS C 1515:2020 class LS/M,class 1/M).
Allows calibration with accuracy of ± 0.10 dB.
Specifications
|
||||||||
Tripod |
|||||||||||
ST-80
![]() This stand can be used for general acoustic measurements. The sound level meter and microphone can be mounted on the stand.
|
Waveform analysis software
AS-70
AS-70
This software allows you to load stored WAVE files from a RION sound level meter, vibration meter or data recorder. Octave, 1/3 octave, and FFT analyses can then be performed. Playback of the real sound files is also possible.
|
||||||||||||||||||||||||||||||||
![]() Frequency analysis screen
|
![]() Frequency analysis screen
|
|||||||||||||||||||||||||||||||
|
Complete software for environmental measurements
Data management software for environmental measurement AS-60
Data management software for environmental measurement AS-60 enables the graph display of measurement data, arithmetic processing, excluded sound processing, preparation of reports, output of files, and playback of real sound files.

Data management screen
- Easy to use
- Reports easy to prepare
- Simultaneous display of
multiple data items(up to 8 data items) - Data on the data recorder
can be loaded(CSV file for DA-40 Viewer) - Data combination

Data management screen
![]() Report preparation screen |
Recommended computer specifications (Common for AS-60/60RT/60VM)
![]() the NX-42EX is also needed. |
Supported models |
|
* Only auto store data are supported.
|
Data management software for environmental measurement AS-60RT
(Includes the octave and 1/3 octave data management software)
(Includes the octave and 1/3 octave data management software)
Adds support for handling octave band analysis data to AS-60
AS-60RT is for managing SX-A1RT, NX-62RT/42RT or NA-28 data on a computer.

Data management screen
Supported models |
|
* Only auto store data are supported.
|
Data management software for environmental measurement AS-60VM
(Includes the vibration level data management software)
(Includes the vibration level data management software)
Adds support for handling data measured with VM-55EX/53A to AS-60
Supported models |
|
* Only auto store data are supported.
|
Types of Measurement
Types of Measurement
Environment | Vehicles | Other industries |
---|---|---|
|
|
|
Specifications
Specifications
NL-62 | ||||
Applicable standards |
IEC 61672-1:2013/2002 class 1 ISO 7196:1995 ANSI/ASA S1.4-2014/Part 1 class 1 JIS C 1509-1:2017 class 1 JIS C 1516:2014 class 1 |
|||
---|---|---|---|---|
CE Marking, WEEE Directives, Chinese RoHS (export model for China only) |
||||
Measurement functions | Simultaneous measurement of the following items, with selected time weighting and frequency weighting | |||
Processing (main ch) |
Instantaneous sound pressure level: Lp Equivalent continuous sound pressure level: Leq Sound exposure level: LE Maximum sound pressure level: Lmax Minimum sound pressure level: Lmin Percentile sound levels: LN (0.1 to 99.9 %, 0.1-increment steps, max. 5 values) |
|||
Processing (sub ch) | Instantaneous sound pressure level: Lp | |||
Additional processing |
One of the following can be selected: C-weighted equivalent continuous sound level: LCeq G-weighted average sound level: LGeq C-weighted peak sound level: LCpeak Z-weighted peak sound level: LZpeak Power average of max. level in time weighted sound level interval LAtm5 I-time-weighted average sound level: LAIeq Max. value of I-time-weighted average sound level: LAImax * Because additional processing frequency characteristics are linked to sub channel frequency characteristics, LAtm5, LAIeq, LAImax can be selected when A characteristics are selected for sub channel. When C, G, or Z characteristics are selected, LCeq and LCpeak, LGeq, and LZ can be selected for additional processing. |
|||
Microphone | Type | UC-59L | ||
Sensitivity level | -27 dB | |||
Measurement range |
A-weighting: 25 dB to 138 dB C-weighting: 33 dB to 138 dB G-weighting: 43 dB to 138 dB Z-weighting: 50 dB to 138 dB C-weighting peak sound level: 60 dB to 141 dB Z-weighting peak sound level: 65 dB to 141 dB |
|||
Inherent noise | A-weighting | 17 dB or less | ||
C-weighting | 25 dB or less | |||
G-weighting | 35 dB or less | |||
Z-weighting | 42 dB or less | |||
Frequency range | 1 Hz to 20 kHz | |||
Frequency weighting | A, C, G and Z | |||
Time weighting | F (Fast) and S (Slow), I (Impulse) and 10 s | |||
Level range | Single range (Linearity range: 113 dB) | |||
Bar graph display range max | Max. 110 dB (20 to 130 dB) | |||
Switching of bar graph display | Set the upper/ lower limit in 10 dB increments. | |||
RMS detection circuit | Digital processing method | |||
Sampling cycle | 20.8 μs (Lp, Leq, LE, Lmax, Lmin, Lpeak : sampling frequency: 48 kHz) 100 ms (LN ) |
|||
Correction functions |
Windscreen correction: Compliant with IEC 61672-1 and JIS C 1509-1 standards when the windscreen is installed. |
|||
Diffuse sound field correction: Correction of frequency characteristics in order to comply with standards (ANSI/ASA S1.4) in diffuse sound field. |
||||
Delay time | The meter can be set to start measuring a specified time (OFF, 1, 3, 5 or 10 s) after the start button has been pressed or when a user-set trigger is exceeded. | |||
Back erase function | When the PAUSE key is pressed to pause measurement, the preceding (user selectable) 0, 1, 3 or 5 s data are excluded from processing. | |||
Display |
Backlit semitransparent color TFT LCD display WQVGA (400 x 240 dots) * LCD with touch panel (Capacitive Touch Panel) |
|||
Numerical display update frequency: 1 s Bar graph update frequency: 100 ms |
||||
Store | Manual | Data for measurement results are stored manually in single address increments. | ||
Number of data |
Internal memory: max. 1 000 sets SD Card: depends on the capacity of the SD Card*1 |
|||
Auto | Instantaneous values (Lp mode) and processed values (Leq mode) are stored continuously and automatically at preset intervals. | |||
Lp sampling cycle | 100 ms, 200 min, 1 s, Leq 1s and user selected time (up to 24 hours) | |||
Leq sampling cycle | 10 s, 1, 5, 10, 15, 30 min, 1, 8, 24 h, and user selected time (up to 24 hours) | |||
Measurement Time | Max. 1 000 h in Auto Lp storage mode, max. 100 000 addresses in Auto Leq storage mode (depends on the capacity of the SD card)*1 | |||
Data recall | Allows viewing of stored data | |||
Setup memory | Up to five setup configurations can be saved in internal memory, for later recall Start up via file settings previously stored on SD card possible | |||
Waveform recording*2 | ||||
File format | Uncompressed waveform WAVE file | |||
Sampling frequency | Select 48 kHz, 24 kHz or 12 kHz | |||
Data length | Select 24 bit or 16 bit | |||
Outputs | DC output | Output DC signals using a frequency weighting characteristic selected by processing. | ||
Output voltage | 2.5 V, 25 mV / dB at bar graph display full scale | |||
AC output | Output AC signal using frequency weighting selected by processing or by A, C, Z, G weighting | |||
Output voltage | 1 V (rms values) at bar graph display full scale | |||
Comparator output |
Turns on when the open-collector output exceeds the set value (max. applied voltage 24 V, max. current 60 mA, allowable dissipation 300 mW). |
|||
USB |
Allows USB to be connected to a computer and recognized as a removable disk Allows USB to be controlled via communication commands |
|||
RS-232C communication | Allows for RS-232C communication via use of a dedicated cable | |||
Data continuous output | ||||
Type of data | Instantaneous value | Lp | ||
Processed value | Leq, Lmax, Lmin, Lpeak | |||
Output interval | 100 ms | |||
Power requirements | Four IEC R6 (size AA) batteries (alkaline or rechargeable batteries) or external power supply | |||
Battery life (23 ℃) | Alkaline battery LR6 (AA): 16 h Ni-MH secondary battery: 16 h At the maximum *Depends on the setting |
|||
AC adapter | NC-98E | |||
External power voltage | 5 to 7 V (rated voltage: 6 V) | |||
Current consumption | Approximately 120 mA (normal operation, rated voltage) | |||
Ambient conditions | Temperature | -10 to +50 ℃ | ||
Humidity | 10 to 90 % RH (non-condensing) | |||
Dustproof / water-resistant performance *3 | IP code: IP54 (except for microphone) See precautions regarding waterproofing |
|||
Dimensions / weight | Approx. 255 (H) x 76 (W) x 33 mm(D), approx. 400 g (with batteries) | |||
Supplied accessories | Storage case x 1, Windscreen WS-10 x 1, Windscreen fall prevention rubber x 1, Hand strap x 1, LR6 (AA) alkaline batteries x 4, SD card 512 MB×1 |
*1 Use Rion fully guaranteed products.
*2 NX-42WR required (sold separately).
*3 Protection against harmful dust and water splashing from any direction.
*2 NX-42WR required (sold separately).
*3 Protection against harmful dust and water splashing from any direction.
Options
Options
Product name | Product number |
---|---|
Waveform recording program (Inst.on 2 GB SD card) | NX-42WR |
Octave, 1/3 octave real-time analysis program (Inst.on 512 MB SD card) | NX-62RT |
Reverberation time measurement program (Inst.on 512 MB SD card) | NX-42RV |
FFT analysis program (Inst.on 512 MB SD card) | NX-42FT |
Data management software for environmental measurement | AS-60 |
Data management software for environmental measurement (Includes the octave and 1/3 octave data management software) |
AS-60RT |
Data management software for environmental measurement (Includes the vibration level data management software) |
AS-60VM |
Waveform analysis software | AS-70 |
SD Card 512 MB | MC-51SD1 |
SD Card 2 GB | MC-20SD2 |
SD Card 32 GB | MC-32SP3 |
AC adapter (100 V to 240 V) | NC-98E |
Battery pack | BP-21A |
Microphone extension cables | EC-04 (from 2 m) |
BNC-Pin output code | CC-24 |
Comparator output cable | CC-42C |
RS 232C serial I/O cable | CC-42R |
USB cable | Generic USB cable can be used |
Sound calibrator | NC-75 |
Pistonphone | NC-72B |
All-weather windscreen | WS-15 |
Windscreen mounting adapter | WS-15006 |
Rain-protection windscreen | WS-16 |
Sound level meter tripod | ST-80 |
All-weather windscreen tripod | ST-81 |
*1 Use Rion fully guaranteed products.
*2 NX-42WR required (sold separately).
*3 Protection against harmful dust and water splashing from any direction.
*2 NX-42WR required (sold separately).
*3 Protection against harmful dust and water splashing from any direction.
[ Precautions regarding waterproofing ]
Before use, verify that the rubber bottom cover and the battery compartment lid are firmly closed.
To maintain the water and dust proof rating, internal packing replacement is required every five years (at cost).
Before use, verify that the rubber bottom cover and the battery compartment lid are firmly closed.
To maintain the water and dust proof rating, internal packing replacement is required every five years (at cost).
Correspondence table for previous model and new model
Correspondence table for previous model and new model(Sound Level Meter class1)
Previous model lineup | New model lineup | ||
---|---|---|---|
|
![]() NL-62 |
Previous model lineup | New model lineup | ||
---|---|---|---|
|
![]() NL-62 + NX-62RT |